DSD/00001/7194.5December 10, 1999* YEAR 2000 READINESS DISCLOSURE STATEMENT *TO OLYMPUS CUSTOMERS PRESENTLY USING:• OLYMPUS DEMAND®, OLYMPUS® AU5000,
DSD/00001/7194.54 *** SYSTEM UP *** # SELECT No.> ---------------------------------------------------
DSD/00001/7194.55 *** COPY FILES *** # SELECT No.> --------------------------------------------------------
DSD/00001/7194.56 CP/M-86 Version 1.0 Segment Address = F000 Last Offset = 317F CP/M86=Version 1.0 A>
DSD/00001/7194.5730. No further action is required.Year 2000 Test Procedure:Use the following steps to test the change to the year 2000 and then reset
DSD/00001/7194.5846. Type 2 and the Enter ↵ key on the keyboard.47. Type 01 for the month and the Enter ↵ key on the keyboard.48. Type 3 and the Enter
DSD/00001/7194.5973. Wait until the DPR displays the (FIG.1) menu.74. Type 7 and the Enter ↵ key on the keyboard.75. Wait until the DPR displays the (
DSD/00001/7194.510 A>GO2000 BACK Transformation 2000 program for OLYMPUS Data ProcessorVer.1.0 Copyright OLYMPUS OPTICAL CO.,LTD All rights reserve
DSD/00001/7194.511 FIG.12Appendix: Error ListError No. Error Message Cause and Corrective Acti
DSD/00001/7194.512Year 2000 Conversion Instructions and Test ProcedureOLYMPUS AU1000 Automated Chemistry AnalyzerYear 2000 Conversion Instructions:1.
DSD/00001/7194.513Year 2000 Conversion Instructions and Test ProcedureOLYMPUS AU5200 Automated Chemistry AnalyzerYear 2000 Conversion Instructions:1.
DSD/00001/7194.52REPLY systems with operating software versions below version 2.4 cannot be upgraded using thespecial software disk.SUPPLEMENTAL INFOR
DSD/00001/7194.514Year 2000 Conversion Instructions and Test ProcedureOLYMPUS AU600 Automated Chemistry AnalyzerYear 2000 Conversion Instructions:12.
DSD/00001/7194.51524. Press the Em. Stop switch to remove power from the analyzer.Press the Reset switch and then the Power switch to restore power an
DSD/00001/7194.516Year 2000 Conversion Instructions and Test ProcedureOLYMPUS PK7200 Automated Microplate SystemYear 2000 Conversion Instructions:14.
DSD/00001/7194.517 _____________________________________________________________ 1 DATE MARK 19
DSD/00001/7194.518 1 DIAGNOSTIC (DPR) 2 DIAGNOSTIC (DPR)
DSD/00001/7194.519 2 DIAGNOSTIC (DPR) 3 DIAGNOSTIC (DPR)
DSD/00001/7194.520Year 2000 Conversion Instructions and Test ProcedureOLYMPUS AU800 Automated Chemistry AnalyzerYear 2000 Conversion Instructions:25.
DSD/00001/7194.53that there are no initialized Data Floppy Disks available upon transition to January 1, 2000, thesystem date can be set to a date pri
DSD/00001/7194.54• Offline Output and Data floppy disk file date stamps will be recorded on the disks as “53” duringthe year 2000. In addition, years
DSD/00001/7194.55° The “Cuvette Status” menu displays a yellow color to alert the operator when 8 or more dayshave elapsed since the last W2/Photocal
DSD/00001/7194.56Director of ServiceOlympus America Inc.Diagnostic Systems DivisionNotice: This Statement contains proprietary information that is pr
DSD/00001/7194.5YEAR 2000 COMPLIANCE LIMITED WARRANTYOFOLYMPUS AMERICA INC.DIAGNOSTIC SYSTEMS DIVISIONEXCEPT AS DESCRIBED IN THE YEAR 2000 READINESS D
DSD/00001/7194.52Limited Warranty, shall not be binding upon OAI-DSD unless reduced to writing and approvedby an expressly authorized officer of Olymp
DSD/00001/7194.53Year 2000 Conversion Instructions and Test ProcedureOLYMPUS REPLYand OLYMPUS AU560 Automated Chemistry AnalyzersYear 2000 Conversion